首頁(yè) > 產(chǎn)品展示 > > 其他檢測(cè)儀器 > CS600高精度碳硫測(cè)定儀

CS600高精度碳硫測(cè)定儀

簡(jiǎn)要描述:簡(jiǎn)單說(shuō)明:CS600提供了一系列*的解決方案檢測(cè)碳和硫。當(dāng)今的燃燒和紅外相結(jié)合的檢測(cè)技術(shù)使您能更準(zhǔn)確、精確、使用方便靈活的剖析各種類型的金屬和無(wú)機(jī)材料。

  • 產(chǎn)品型號(hào):CS600
  • 廠商性質(zhì):生產(chǎn)廠家
  • 更新時(shí)間:2024-10-08
  • 訪  問(wèn)  量:1743
詳情介紹

詳細(xì)說(shuō)明:

CS600提供了一系列*的解決方案檢測(cè)碳和硫。當(dāng)今的燃燒和紅外相結(jié)合的檢測(cè)技術(shù)使您能更準(zhǔn)確、精確、使用方便靈活的剖析各種類型的金屬和無(wú)機(jī)材料。

Carbon and Sulfur in Steel, Nickel-Base, and Cobalt-Base Alloys
 
Instrument
CS600-Series
 
Sampling and Sample Preparation
Surface contamination on the sample can cause significant errors in the analytical data; therefore, care must be taken to ensure a clean, representative sample is analyzed. Solid
samples should be abraded with a clean file, rinsed in acetone and dried with warm air prior to analysis. Samples that cannot be abraded due to irregular shapes should be rinsed in a
suitable solvent such as acetone, and dried with warm air. Care
must be taken to remove all traces of the solvent. If a sample is porous, refrain from using solvents,
as it will be difficult to remove all traces of the solvent by drying. Refer to ASTM E1806 for additional sampling and sample preparation information.
 
Accessories
528-018 or 528-018HP Ceramic Crucibles (preheated)*; LECOCEL II (501-008) or
LECOCEL II HP (502-173) accelerator.
 
*Ceramic crucibles are baked in a muffle or tube furnace (LECO TF-10) at 1250oC for a minimum of
15 minutes, or at 1000oC for 40 minutes. The crucibles are removed from the furnace, allowed to cool for

 
 
 
 
1 to 2 minutes, and are transferred to a desiccator for storage. If the crucibles are not used within four hours, they should be re-baked.

 
Calibration
There are several suitable calibration samples available from LECO. Likewise, NIST, JK, JSS, and BCS are certified bodies that have a variety of certified reference materials (SRM/CRM) available as well. It is important that both high- and low-carbon ranges are calibrated. Single or multipoint calibration curves
can be utilized. Refer to the operator's instruction manual for details.
 
Method Parameters
Purge Time (seconds):
1**
 
Delay Time (seconds):
8**
 
Furnace Low Power (%):
100
 
Furnace High Power (%):
100
 
Furnace Ramp Rate:
0
 
 
 
Carbon
 
Sulfur
Minimum Timeout (seconds):
40
40
Comparator Level:
1.00
1.00
Significant Digits:
4 or 5
4 or 5
Integration Delay:
0
0
 
 


 

**For optimum analytical performance on samples with low carbon and
sulfur content 0.05%), a purge time of 10 seconds and a delay time of
20 seconds is recommended.

 
 
CS600


 

Procedure
1.   Prepare instrument for operation as outlined in the operator's instruction manual.
2.   Determine blank.
a.     Enter 1.0000 g weight into weight stack. b.      Add ~1.5 g of accelerator to crucible.
c.      Place the crucible on the furnace pedestal (or appropriate autoloader position if so equipped), and analyze.
d.     Repeat steps 2a through 2c a minimum of five times.
e.     Enter blank following procedure outlined in the operator's instruction manual.
3.   Calibrate.
a.     Weigh ~1.0 g calibration sample into crucible and enter weight into weight stack.
b.     Add ~1.5 g of accelerator on top of sample.
c.      Place the crucible on the furnace pedestal (or appropriate autoloader position if so equipped), and analyze.
d.     Repeat steps 3a through 3c a minimum of five times for each calibration sample intended for calibration.
e.     Calibrate using the procedure outlined in the operator's instruction manual.
4.   Analyze Samples.
a.     Weigh ~1.0 g sample into crucible and enter weight into weight stack.
b.     Add ~1.5 g of accelerator on top of sample.
c.      Place crucible on furnace pedestal (or appropriate autoloader position if so equipped), and analyze.
 
Note:   Some "solid" samples, notably nickel and/or cobalt-base alloys, are difficult to combust and may require additional accelerator. Likewise, reducing the sample to small chunks, chips, or pieces will facilitate combustion. For additional information on Carbon and Sulfur Determination using an Induction Furnace IR detection instrument, request LECO's Application Tips CD (209-050-055).
 
Typical Results

Sample
Weight (g)
C %
S %
Sample
Weight (g)
C %
S %
LECO
1.0
0.605
0.0176
LECO
1.0
0.00089
0.00062
501-505
 
0.605
0.0172
502-348
 
0.00089
0.00067
Steel Ring
 
0.605
0.0169
Low C and S
 
0.00094
0.00068
0.607 % C
 
0.607
0.0171
Steel Pin
 
0.00095
0.00060
0.0172% S
 
0.607
0.0174
0.0009 % C
 
0.00085
0.00064
 
 
0.605
0.0176
0.0007 % S
 
0.00085
0.00061
 
 
0.606
0.0178
 
 
0.00090
0.00062
 
 
0.604
0.0173
 
 
0.00090
0.00060
 
 
0.606
0.0173
 
 
0.00090
0.00065
 
 
0.605
0.0168
 
 
0.00089
0.00055
 
X =
0.606
0.0173
 
X =
0.00090
0.00062
 
s =
0.001
0.0003
 
s =
0.00003
0.00004
 
LECO
 
1.0
 
0.0375
 
0.0223
 
LECO
 
1.0
 
0.0414
 
0.0025
501-501
 
0.0380
0.0221
501-933
 
0.0414
0.0027
Steel Ring
 
0.0375
0.0222
Ni-Base Alloy
 
0.0413
0.0027
0.0377 % C
 
0.0377
0.0222
Chip Sample
 
0.0414
0.0028
0.0220 % S
 
0.0376
0.0223
0.042 % C
 
0.0415
0.0027
 
 
0.0376
0.0222
0.003 % S
 
0.0413
0.0028
 
 
0.0373
0.0221
 
 
0.0414
0.0028
 
 
0.0376
0.0222
 
 
0.0415
0.0028
 
 
0.0377
0.0223
 
 
0.0412
0.0027
 
 
0.0377
0.0222
 
 
0.0414
0.0029
 
X =
0.0376
0.0222
 
X =
0.0414
0.0027
 
s =
0.0002
0.0001
 
s =
0.0001
0.0001


 

留言框

  • 產(chǎn)品:

  • 您的單位:

  • 您的姓名:

  • 聯(lián)系電話:

  • 常用郵箱:

  • 省份:

  • 詳細(xì)地址:

  • 補(bǔ)充說(shuō)明:

  • 驗(yàn)證碼:

    請(qǐng)輸入計(jì)算結(jié)果(填寫(xiě)阿拉伯?dāng)?shù)字),如:三加四=7
技術(shù)支持:化工儀器網(wǎng)   sitemap.xml   管理登陸
©2024 版權(quán)所有:寧波市鄞州瑾瑞儀器設(shè)備有限公司   備案號(hào):浙ICP備14028217號(hào)-3

浙公網(wǎng)安備33020402000311號(hào)

望奎县| 宜君县| 盘锦市| 颍上县| 河津市| 乐昌市| 平和县| 临邑县| 平遥县| 鄂州市| 乐平市| 元朗区| 广州市| 丰宁| 莒南县| 武乡县| 亳州市| 绵阳市| 调兵山市| 慈利县| 宣城市| 永定县| 类乌齐县| 呈贡县| 策勒县| 阿瓦提县| 易门县| 盐亭县| 延长县| 汪清县| 洪江市| 宿迁市| 营山县| 勐海县| 克拉玛依市| 乐山市| 建德市| 突泉县| 察雅县| 晋城| 许昌县|